User manual
ABPM-05, ABPM-04 and BlueBP-05
ambulatory blood pressure monitors with
CardioVisions software
(from version 1.18)
Vista de pagina 0
1 2 3 4 5 6 ... 33 34

Indice de contenidos

Pagina 1 - User manual

User manual ABPM-05, ABPM-04 and BlueBP-05 ambulatory blood pressure monitors with CardioVisions software (from version 1.18)

Pagina 2 - Table of contents

10 Information displayed on ABPM-04 LCD screen only: Rectangle blinking: measurement is progress (mmHg) Battery voltage display (5,37V) Com

Pagina 3 - YYYY/nnnnnn

11 6. Technical parameters Technical parameters ABPM-05 BlueBP-05 ABPM-04 Power supply 2 AA rechargeable NiCd or NiMH batteries or 2 AA alkaine bat

Pagina 4

12 Please note that the device may not meet its performance specifications if stored or used outside the specified environmental conditions. Bot

Pagina 5 - 2. Devices

13 8. Safety concerns Electric shock hazard protection Each ABPM recorder meets the relevant shock hazard protection standards. ABPM-05 and BlueBP-0

Pagina 6 - 3. Accessories

14 BlueBP-05 is capable of wireless communication with the computer using Bluetooth technology. For successful communication, you need

Pagina 7 - 4. Using the buttons

15 If you wish to use a BlueBP-05 with Bluetooth under Windows XP SP 2, follow the steps below. With other Windows versions, please refer to the docu

Pagina 8 - 5. Display

16 8. Apply the cuff to the patient with the device placed in the pouch. 9. Give the patient diary to the patient along with detailed instructions ab

Pagina 9

17 10 second delay state 3. measurement plan: 30/30 minute day/night intervals 1. measurement plan: 15/30 minute day/night intervals Progra

Pagina 10

18 10. Remove the unit and the cuff from the returned patient. 11. Ask for the patient diary and ask the patient for any events, symptoms, observatio

Pagina 11 - 6. Technical parameters

19 In order to change batteries, take the recorder out of the holder pouch and remove the battery compartment cover on the back-side.

Pagina 12 - 7. Care and maintenance

2 Table of contents 1.Recommended use of ambulatory blood pressure monitors ... 5 2.Devices ...

Pagina 13 - 9. The use of the monitors

20 12. Using CardioVisions software The CardioVisions software package provides means to handle Meditech ambulatory blood pressure and ECG recorder

Pagina 14

21 Suggested CPU: 2 GHz P4 or equivalent RAM: 512 MB VGA: 1024x768 @ 16 bit color LAN: 100 Mbit/s HDD: about 40 MB for the software + disk space for

Pagina 15 - Monitoring step by step

22 - Welcome screen - System Administrator settings Your default login name and password is admin/admin. Default settings can be mo

Pagina 16

23 Here you are required to create a user account by filling in the empty spaces, but at least the bold entries. Creating a user is necessary to see

Pagina 17

24 Installing CardioVisions Network Edition If you wish to access CardioVisions functions from several workstations in a computer network

Pagina 18 - 10. Batteries

25 Backup Creating a backup from CardioVisions database is strongly recommended to prevent data loss. It is advised to create a backup often, mo

Pagina 19

26 3.) After clicking ’Tools/Archive examinations from the selected folder’, a new panel will appear on your screen, which will inform y

Pagina 20

27 database. To open the examination like this, the program will ask the destination route of the archived data. If you have saved the data on the ha

Pagina 21

28 13. Meditech product warranty information (a) RECORDER WARRANTY. The main recorder unit will be free from defects in materials and

Pagina 22

29 14. CardioVisions software license agreement Important – read carefully: This License Agreement is a legal agreement between you (e

Pagina 23

3 Important information on ABPM-05, ABPM-04 and BlueBP-05 systems. Read carefully, please. This symbol on a Meditech recorder is a

Pagina 24

30 Miscallaneous This Agreement is governed by the laws of Hungary. Should you have any questions concerning this Agreement, please cont

Pagina 25 - Archiving

31 15. EMC information Medical electrical equipment should be used with precautions according to EMC, and must be installed according to the EMC n

Pagina 26

32 Electromagnetic immunity Meditech ambulatory blood pressure monitors are suitable for use in the specified electromagnetic environment

Pagina 27

33 ABPM-05 Power frequency (50/60Hz) magnetic field IEC 61000-4-8 3A/m 3A/m Power frequency magnetic fields are at levels characteristic of a typica

Pagina 28

34 BlueBP-05 Radiated RF IEC 61000-4-3 3V/m 80MHz-2,5GHz 3V/m Recommended Separation distance: d=[3,5/V1]√P d=[3,5/3V/m]√P (80MHz – 800MHz) d=[7/3V

Pagina 29 - Copyright

4 Blood pressure measurements determined with the algorithm of an ABPM-05, ABPM-04 or a BlueBP-05 recorder on adults are equivalent to tho

Pagina 30

5 1. Recommended use of ambulatory blood pressure monitors Indications for ambulatory blood pressure monitoring The following indications are li

Pagina 31 - 15. EMC information

6 Start button LCD Interface connector Event button On the front side of the housing, there is the LCD, the buttons of the device and a sticker disp

Pagina 32 - Electromagnetic immunity

7 4. Using the buttons On the front side of ABPM-05 and BlueBP-05, below the LCD, you can find three buttons: START button is marked by a tri

Pagina 33

8 Marking the time of going to bed and rising from bed – ABPM-05and BlueBP-05 only If the DAY/NIGHT shift function is disabled during programming, th

Pagina 34

9 Systolic value of just completed measurement (mmHg) The blood pressure measurement is cancelled by pressing a button. Diastolic value of ju